G-arm medical imaging
G-arm medical imaging | |
---|---|
Purpose | fluoroscopic X-ray for diagnosis |
G-arm medical imaging systems are based on fluoroscopic X-ray and are used for a variety of diagnostic imaging and minimally invasive surgical procedures. The name is derived from the G-shaped arm used to connect two X-ray generators and two X-ray detectors, image intensifiers or digital flat panel detectors, to one another. The main advantage of the G-arm, compared to a conventional C-arm system, is that it combines a pair of X-ray chains facilitating simultaneous views in two perpendicular planes, also called G-arm imaging.
Background
[edit]Fluoroscopic X-ray is used for a variety of diagnostic imaging and minimally invasive surgical procedures. For surgery it is practical to use a mobile C-arm system where the x-ray generator and x-ray detector are placed on a C-shaped arm positioned directly opposite from and aligned centrally to each other. The C-arm can be moved horizontally, vertically and around the swivel axes, so that X-ray images of the patient can be produced from different angles. The disadvantage of the C-arm is that imaging is limited to only one plane at a time.
From C-arm to G-arm
[edit]For precision surgery procedures, for example in orthopedics,[1] it is necessary to generate a view of the surgical site from two perpendicular viewing positions, usually a frontal and a lateral view. With a C-arm system this can only be achieved by manually repositioning the C-arm for the second view. This can be challenging due to the presence of sterile drapes and other surgical devices. Bringing the x-ray generator from a non-sterile area, underneath the operating table to a horizontal position close do the patient can compromise the sterility of the surgical area and increase the risks of infections. Shifting the C-arm must normally be done many times during the surgical procedure, taking additional time and increasing the radiation exposure to patient and staff. With the G-arm the stand can stay in a fixed position during the entire surgical procedure as both imaging planes can be viewed simultaneously. Benefits are shorter procedures,[2][3] higher accuracy,[4][5][6] and lower radiation exposure.[7][8][9] The fixed position of the stand also ensures a better sterility.
Technical description
[edit]A G-arm medical imaging system is typically composed by the following basic elements:
A. A movable stand
B. X-ray source and detector for the frontal view.
C. X-ray source and detector for the lateral view
D. Patient table with extensions (here shown in position for a hip surgery)
E. Control unit with post-processing image software[10] and displays for the two perpendicular imaging planes
X-ray detectors, the height of the stand and the angle of the G-Arm may be motorized or manually moved for fine adjustment of the iso-center of the x-ray beams.
Clinical applications
[edit]Due to the benefits resulting from the ability to simultaneously view two perpendicular planes G-arm Medical Imaging systems are mainly used for orthopedic and traumatic surgery procedures, such as hips, knees and spine.
History
[edit]The G-Arm Imaging system is originating from a patented invention[11] from 1970 by the Swedish company Saab AB, an aerospace and automobile manufacturer who at that time also was active within the field of medical devices. In the original product the two x-ray chains were placed in an annular stand. The product, called Multiplane, had some practical disadvantages due to the closed form of the stand. Later a part of the annular stand was cut out and the G-arm was born.
The G-Arm concept has been gradually developed to its current form. In 2002 the concept was acquired by the company Scanflex Healthcare AB based in Stockholm, Sweden who has refined the product and is marketing it worldwide under the trademark, Biplanar (trademark registration cancelled 2011).[12] The G-Arm concept was also adopted in 2011 by Beijing East Whale Imaging Tech Co., who hold the registered trademarks for G-Arm and MultiScan G-Arm System under this category[13][14] and a number of related patents.[15] The company was founded in 1998 and is based in Beijing, China. Whale Imaging Inc, its R&D and marketing headquarters, is based near Boston, USA.[16] The Whale G-Arm received a Spine Technology award in 2014.[17]
References
[edit]- ^ Stern, Steven (2000-12-13). Key Techniques in Orthopaedic Surgery (1 ed.). Thieme Medical Publishers. ISBN 9783131248510.
- ^ kopenmac09 (2015-03-10), Technique for Spinal Injections supported by Biplanar 400E - Swemac Medical Appliances, retrieved 2016-03-18
{{citation}}
: CS1 maint: numeric names: authors list (link) - ^ Instructional Course Lectures, Vol. 54. American Academy of Orthopaedic Surgeons. 2005-01-31. ISBN 9780892033454.
- ^ Westberry, David E.; Davids, Jon R.; Cross, Andrew; Tanner, Stephanie L.; Blackhurst, Dawn W. (2008-02-01). "Simultaneous biplanar fluoroscopy for the surgical treatment of slipped capital femoral epiphysis". Journal of Pediatric Orthopedics. 28 (1): 43–48. doi:10.1097/BPO.0b013e3181558bee. ISSN 0271-6798. PMID 18157045. S2CID 28878176.
- ^ Killian, J. T.; Conklin, M. J.; Kramer, T.; White, S. (1999-01-01). "Improved percutaneous slipped capital femoral epiphysis treatment: continuous biplanar fluoroscopy and proper guide wire selection". Journal of the Southern Orthopaedic Association. 8 (3): 155–162, discussion 162. ISSN 1059-1052. PMID 12132860.
- ^ Strömqvist, B.; Hansson, L. I.; Nilsson, L. T.; Thorngren, K. G. (1984-10-01). "Two-year follow-up of femoral neck fractures. Comparison of osteosynthesis methods". Acta Orthopaedica Scandinavica. 55 (5): 521–525. doi:10.3109/17453678408992951. ISSN 0001-6470. PMID 6507073.
- ^ Cason, Garrick; Bonvallet, Todd; Rouben, David; Rudd, Joseph; Ervin, Timothy (2008). "40. Radiation Exposure to Operating Room Personnel During Minimally Invasive Spine Surgery: A Comparison of Single vs. Simultaneous Dual Fluoroscopy". The Spine Journal. 8 (5): 20S. doi:10.1016/j.spinee.2008.06.046. Retrieved 2016-03-18.
- ^ Romanova, K.; Vassileva, J.; Alyakov, M. (2015-07-01). "Radiation exposure to the eye lens of orthopaedic surgeons during various orthopaedic procedures". Radiation Protection Dosimetry. 165 (1–4): 310–313. doi:10.1093/rpd/ncv122. ISSN 1742-3406. PMID 25944961.
- ^ Fan, Guoxin; Fu, Qingsong; Gu, Guangfei; Zhang, Hailong; Guan, Xiaofei; Zhang, Lei; Gu, Xin; He, Shisheng (2015). "Radiation Exposure to Surgeon in Minimally Invasive Transfor... : Clinical Spine Surgery". Journal of Spinal Disorders and Techniques. 28 (3): E173–E180. doi:10.1097/BSD.0000000000000210. PMID 25353207. S2CID 5362757.
- ^ "UPPSATSER.SE: Enhancement of X-ray Fluoroscopy Image Sequences using Temporal Recursive Filtering and Motion Compensation". www.uppsatser.se. Retrieved 2016-03-18.
- ^ "US3549885.pdf" (PDF). docs.google.com. Retrieved 2016-03-18.
- ^ "BIPLANAR Trademark of Scanflex Healthcare AB Serial Number: 79099559 :: Trademarkia Trademarks".
- ^ "G-ARM Trademark of BEIJING EAST WHALE IMAGE TECHNOLOGY CO., LTD. Serial Number: 85664902 :: Trademarkia Trademarks".
- ^ "MultiScan G-arm System". European Union Intellectual Property Office (EUIPO).
- ^ "Google".
- ^ "Company | Whale Imaging".
- ^ "Eleven Spine Technologies Won "Best New Technology for 2014" | 10th Annual Spine Technology Awards".