Jump to content

Autocorrelator

From Wikipedia, the free encyclopedia

A real time interferometric autocorrelator is an electronic tool used to examine the autocorrelation of, among other things, optical beam intensity and spectral components through examination of variable beam path differences. See Optical autocorrelation.

Description

[edit]

In an interferometric autocorrelator, the input beam is split into a fixed path beam and a variable path beam using a standard beamsplitter. The fixed path beam travels a known and constant distance, whereas the variable path beam has its path length changed via rotating mirrors or other path changing mechanisms. At the end of the two paths, the beams are ideally parallel, but slightly separated, and using a correctly positioned lens, the two beams are crossed inside a second-harmonic generating (SHG) crystal. The autocorrelation term of the output is then passed into a photomultiplying tube (PMT) and measured.

Details

[edit]

Considering the input beam as a single pulse with envelope , the constant fixed path distance as , and the variable path distance as a function of time , the input to the SHG can be viewed as

This comes from being the speed of light and being the time for the beam to travel the given path. In general, SHG produces output proportional to the square of the input, which in this case is

The first two terms are based only on the fixed and variable paths respectively, but the third term is based on the difference between them, as is evident in

The PMT used is assumed to be much slower than the envelope function , so it effectively integrates the incoming signal

Since both the fixed path and variable path terms are not dependent on each other, they would constitute a background "noise" in examination of the autocorrelation term and would ideally be removed first. This can be accomplished by examining the momentum vectors

If the fixed and variable momentum vectors are assumed to be of approximately equal magnitude, the second harmonic momentum vector will fall geometrically between them. Assuming enough space is given in the component setup, the PMT could be fitted with a slit to decrease the effect the divergent fixed and variable beams have on the autocorrelation measurement, without losing much of the autocorrelation term. can then be assumed to be nearly equal to

which gives the autocorrelation as a function of , the difference in path lengths.

Technical specifications

[edit]
  • Calibration Factor -- the factor to convert real-time to pulse delay time when viewing the output of the autocorrelator. One example of this would be 30 ps/ms in the Coherent Model FR-103 scanning autocorrelator, which suggests that a 30 ps pulse autocorrelation width would produce a 1 ms FWHM trace when viewed on an oscilloscope.
  • Time Resolution -- related to the time constant of the PMT, an estimate can be made by multiplying the time constant with the calibration factor.

See also

[edit]

References

[edit]
  • Coherent Mira Model 900 Laser Operator's Manual
  • Principles of Communications: Systems Modulation and Noise 5ed, Rodger E. Ziemer and William H. Tranter
  • Elements of Optoelectronics and Fiber Optics, Chin-Lin Chen